6.8.6. Volume
Configure volume calculation method.
For details information of Tank configuration refers
Table 9.
The Level Transmitter measures only distance and related quantities (level, percent of range, etc.). The
calculation of volume by the transmitter is based on measured level and additional tank geometry
measurements.
Wet Volume Calibration
When the measured level for the tank reaches a level where the corresponding volume is known, select
a row in the method, enter the corresponding volume in the textbox below, and proceed further. This
will immediately set the values into the selected row in the strapping table
(I.e. immediately set in the transmitter).