Table 5-9 Changing Security Level - Honeywell 7866 Operation And Maintenance Manual

Triple range digital h2 and co2 analyzer/indicator and gas sampling panel
Table of Contents

Advertisement

Changing Security Level
If the LOCKOUT feature has been set to anything other than NONE and a security code has been entered,
you will have to change the LOCKOUT level to NONE before calibrating.
Step
Press
1
SET UP
2
FUNCTION
3
or
FUNCTION
4
or
5
FUNCTION
6
9/06
7866 Triple Range Digital H

Table 5-9 Changing Security Level

until you see:
Upper Display:
SETUP
Lower Display:
UNIT
You will see SECURITY in the Lower Display.
to enter the four-digit security code number (0001 to 9999)
If the security code is unknown, see Appendix B – Security Bypass.
You will see LOCKOUT in the Lower Display.
Select NONE.
Saves selection to nonvolatile memory.
and CO
Analyzer/Indicator - Operation and Maintenance
2
2
Action
Calibration