Instrument Test Procedures For The Vibrex 2000 And Vibrex 2000 Plus - Honeywell Vibrex 2000 User Manual

Digital dynamic balancer / analyzer systems
Table of Contents

Advertisement

6.3.5.2
The following tests provide the highest confidence that the
instrument is accurately measuring the incoming sensor signals.
6.3.5.3
Signals from a calibrated source will be measure at start, mid,
and end band frequencies of the filters used and compared to
insure they meet specifications. In addition, a mid band signal
will be measured for amplitude, phase, and frequency response
and compared to specification.
All the above testing will be repeated for each of the 2 signal
paths.
Test Equipment:
HP 8904A
Signal
Generator
BNC to Velo
cable 8278
BNC to Mag
cable 10502
Fluke True
RMS meter
Honeywell Confidential: Use or disclosure of information on this page is subject to the restrictions on the title page.
Instrument Test Procedures for the
Vibrex 2000 and Vibrex 2000 Plus
Approach:
Hewlett-Packard
Honeywell
Honeywell
Fluke
13841 Rev A
Chapter 6 – Maintenance
10 mV to 5 VAC (+/-0.1%),
distortion plus noise less
than -60 dB, sync to sine
signal phase relation
ship180 degrees
5% 2 Hz - 10 kHz
219

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vibrex 2000 plusEnhanced vibrex 2000 plus

Table of Contents