Chapter 5 – General Operation
Honeywell Confidential: Use or disclosure of information on this page is subject to the restrictions on the title page.
Use the
context key to change measurement
MEAS
name. Some Regimes may have only one measurement
so pressing this key will have no effect.
CAUTION: A common mistake when first using the
EV2K+ is to select the wrong measurement before
starting an acquisition. Since the measurement is
directly tied to the internal chart used in the solution,
selecting the wrong measurement type for the input
channel will give the wrong solution. For example, you
do not want to do a "M/R L" (lateral) solution on the
vertical sensor. The EV2K+ usually defaults to lateral
on CH 1A and vertical on CH 2B, but the MEAS and
BAL CHAN can be changed by the user.
A default
BAL CHAN
measurement name but the selection can be changed by
toggling through all available channel combinations.
CAUTION: Make sure the physical EV2K+ connection
is the same one you use for this selection. The EV2K+
allows for two additional vibration channels (3 and 4)
using an optional cable adaptor.
Use the
AutoCorrection
and OFF. The default is OFF. Since the factory charts
are specifically setup for a given condition, it is
recommended that this option not be used unless balance
cannot be achieved in an acceptable number of runs.
With proper selections, press the
is selected corresponding to the
key to turn auto-correction ON
START
13841 Rev A
166
to continue.