Replace speaker assembly

7-42
Field replaceable units: Replace main unit parts
4. Remove the speaker harness and the optional alarm light cable
harness from the top of the writer board.
5. Remove three screws holding the writer flex to the writer bracket.
6. Replace writer assembly and/or writer flex and reassemble the
patient monitor in reverse order.
7. Complete the procedures in
1. Remove the DAS assembly according to steps starting on page 7-37.
2. Remove the writer assembly/writer flex according to steps on
page 7-41.
3. Remove the writer bracket from the frame by squeezing the sides
together.
4. Disconnect the speaker cable connector from the processor board.
Dash 3000/4000/5000
"Recommended checkout"
on page 7-52.
532A
586B
2000966-542D