No waveform at central station

6-10
Troubleshooting: No waveform at central station
If you do not have a waveform at the central station:
Ensure the central station software is compatible.
Check all cables for a good connection.
Check the configured alarms and manual print locations at the
patient monitor.
Ensure the care unit name is the same in the patient monitor and in
the central station.
Ensure the internet protocol (IP) addresses are configured correctly.
For a troubleshooting tree with more information, refer to
waveforms or parameters are displayed at the CIC Pro center"
page B-5.