Service Mode options
4-6
Configuration: Service menus
Following is a list of options in the main code service menu.
Menu option
REVIEW ERRORS
CALIBRATE
BATTERY SERVICE
NOTE
If an INCOMPAT message displays in the Manufacturer Name
column and UNKNOWN displays in all other columns, the
installed battery is incompatible.
PATIENT-MONITOR TYPE
WARNING
Changing the patient-monitor type will default the
admit function to Standard configuration. Different
alarms and parameters are activated for each selection.
NOTE
The keypad/remote control is DIDCA programmed for specific
patient monitor types. The error message, "WARNING:
REMOTE MISMATCHED WITH MONITORING MODE"
displays if the patient monitor and keypad/remote control do not
match.
Dash 3000/4000/5000
Service Mode menu options
Description
For advanced troubleshooting by GE engineers. Error
log data can be transferred over the network to a central
station and then loaded onto a diskette for review. Refer
to
"Error messages"
on page 6-5 for details.
For checkout or calibration of the noninvasive blood
pressure, ECG analog output, BP analog output, CO
service, and SAM service menu functions of the patient
monitor.
Provides battery data.
For selecting the type of patient monitor desired, i.e
adult, neonatal or operating room. Refer to
Monitor Type"
on page 4-12 for detailed procedures.
2
"Patient-
2000966-542D