Pulse oximetry tests for Nellcor OxiMax SPO2
Equipment
Procedures
2000966-542D
Functional and electrical safety checks: Functional Checkout procedures
Use the following equipment for these tests:
Nellcor SRC-MAX SpO
NOTE
The Nellcor SRC-MAX SPO
the Nellcor SPO
1. Verify that 2 AA alkaline batteries are installed in the SRC-MAX
Nellcor Pulse Oximetry functional tester.
2. Connect the SRC-MAX Nellcor Pulse Oximetry functional tester to
the 9-pin end of sensor connector from the Dash. Make sure the Dash
is powered.
3. Verify that the IR LED and RED LED drive indicators are both lit on
the SRC-MAX.
4. Verify the SRC-MAX default indicators are set as follows:
HEART RATE = 60 bpm
LIGHT = LOW
%SpO2 = 75
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.