GE Dash 3000 Service Manual page 213

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Acceptance criteria NC
Acceptance criteria SFC – ground (earth), line or neutral open
2000966-542D
Functional and electrical safety checks: Electrical safety tests
Neutral – CLOSED
GND (Earth) – OPEN
8. Read and record the current leakage indicated on the tester.
9. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – OPEN
10. Read and record the current leakage indicated on the tester.
11. Change leakage tester switches to:
Polarity – REVERSED
Neutral – OPEN
GND (Earth) – CLOSED
12. Read and record the current leakage indicated on the tester.
13. Change leakage tester switches to:
Polarity – REVERSED
Neutral – CLOSED
GND (Earth) – CLOSED
14. Read and record the current leakage indicated on the tester.
15. Set the power switch of the device to OFF.
16. Repeat the steps in this procedure using the appropriate SPO2 Test
Body. Connect the SPO2 Test Body to the blue SPO2 connector of the
device under test.
With Ground and Neutral CLOSED – If reading is greater than 10 µA,
the device under test fails. Contact GE Technical Support
If any reading is greater than 50 µA, the device under test fails. Contact
GE Technical Support.
Dash 3000/4000/5000
8-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000Dash 5000001c 051d 003a

Table of Contents