Open display assembly

Dash 3000 patient monitor
with display shield flex circuit
Separate the patient monitor bezel from
the LCD display.
1. Remove and retain the four screws
holding the display shield to the LCD
display and two screws holding the
display shield to the patient monitor
bezel.
4 larger screws
2 smaller screws
2. Remove the display assembly from the
patient monitor bezel.
3. Pull the LCD display isolator out of the
way. Disconnect the LCD display cable.
LCD display cable
4. Disconnect the cable from the backlight
inverter PCB.
cable
2000966-542D
Field replaceable units: Replace display assembly parts
Follow the steps below for the patient monitor you are servicing.
Dash 4000 patient monitor
1. Remove and retain the two screws from
the display flex cable. Pull up on the
latch tabs to remove the flex cable from
the display.
2 screws
2. Remove and retain the 10 screws as
shown.
10 screws
500A
3. Separate the transition bezel from the
front bezel. It is not necessary to remove
501A
the metal backplate.
4. Set aside the transition bezel.
NOTE
The display that is pictured above is a
Sharp.
502A
Dash 3000/4000/5000
Dash 5000 patient monitor
1. Remove and retain the two screws from
the display flex cable. Pull up on the
latch tabs to remove the flex cable from
the display.
2 screws
879A
2. Remove and retain the 14 screws as
shown.
14 screws
899A
3. Separate the transition bezel from the
front bezel. It is not necessary to remove
the metal backplate.
4. Set aside the transition bezel.
910A
860A
900A
884A
7-27