Fault Analysis; Overview; Required Tools Or Equipment; Problems - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Fault analysis

Overview

Required tools or equipment

Problems

Problem
General
The unit is plugged in, but it does not
switch to AC from battery power.
Alarm audio
Audio does not sound.
Go to the Alarm Help Menu and select
"Crisis." Listen for the three-beep
audio.
Waveform display
6-2
Troubleshooting: Fault analysis
The symptoms and solutions in this chapter represent only a few of the
faults that you may encounter and are not intended to cover every
possible problem that may occur.
A systematic approach to the diagnosis of problems as well as a general
understanding of the architecture, both hardware and software, of the
patient monitor are essential to ensure successful troubleshooting of a
device. GE recommends formal service training before repairs are
attempted. These troubleshooting procedures combined with training
provide the service technician with skills necessary to service and repair
a patient monitor in the event of a malfunction.
Standard set of hand tools
Patient simulator with cables
Digital multimeter
Before starting any detailed troubleshooting, complete a thorough visual
inspection of the following.
All cable connections secure?
Devices properly powered?
Connected to a proper power source?
Possible reason
The output voltage of the power
supply is inadequate for the
processor PCB to recognize.
The speaker cable is loose or
disconnected.
The speaker failed.
The audio circuit on the processor
PCB failed.
Patient cables are bad.
The acquisition PCB, DAS Assembly
or power supply PCB failed.
Dash 3000/4000/5000
Solution
Perform
"Power outlet test" on page 8-5
AC power.
Replace the power supply.
Replace the processor/power management
assembly.
Restore the cable connections.
Replace the speaker.
Replace the Replace the processor/power
management assembly.
Perform appropriate sections of the
Checkout procedures" on page 8-20
the waveform problem.