Optional components

TRAM-rac 2A module housing
Dash Port 2 docking station
2-10
Equipment overview: Components
Verify compatibility of an unmarked battery as follows.
1. Install a battery pack in the patient monitor.
2. Using the Trim Knob control, access the Service Mode menu
starting from the Main Menu. Select MORE MENUS > MONITOR
SETUP > SERVICE MODE.
3. Enter password using the Trim Knob control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
4. Select BATTERY SERVICE.
5. Verify that the MANUFACTURER NAME does not display
INCOMPAT, NME, or UNKNOWN for the battery corresponding to
BATTERY A or BATTERY B slot.
The TRAM-rac 2A module housing currently supports the SAM and ICG
modules.
An integral power supply is used to run the TRAM-rac 2A and support
the needed voltages.
The docking station is a quick mount/dismount base for a Dash patient
monitor. It gives the patient monitor easy connect or disconnect access to
AC power, CARESCAPE Network, a remote display, and auxiliary
devices.
See the Dash Port 2 Docking Station Operating Instructions and the
Dash Port 2 Docking Station Service Manual for additional information.
NOTE
When a Dash patient monitor is connected to the docking station,
only the docking station's Ethernet port is active. The Dash patient
monitor's network port remains inactive until the patient monitor is
disconnected from the docking station.