Select the monitoring device
Select the error log date
6-22
Troubleshooting: Get error logs
4. Select the Care Unit.
5. Select UNIT:
6. Using the Trim Knob control, change the displayed Care Unit
name. When the desired Care Unit name displays, press the Trim
Knob control.
1. Select DEVICE:
2. Using the Trim Knob control, change the displayed device name.
Note that only monitoring devices within the previously selected
Care Unit show. When the desired monitoring device name displays,
press the Trim Knob control.
1. Select DATE:
2. Using the Trim Knob control, change the error log date. Note that
one of the selections is ALL, which retrieves all stored error logs from
the specified device. When the desired date displays, press the Trim
Knob control.
Dash 3000/4000/5000
048A
044A
045A
046A
2000966-542D