Pulse oximetry tests for GE Ohmeda SPO2 oximeter

Equipment
Procedure
8-30
Functional and electrical safety checks: Functional Checkout procedures
3. Change the patient simulator BP output to 240 mmHg.
4. Observe a reading of 240/240 (240) ± 4 mmHg in the BP parameter
window on the patient monitor display.
5. Change the patient simulator BP output to WAVE or to a 120/80
wave form (simulated BP waveform).
6. Observe a distortion-free BP waveform and a reading of
approximately 120/80 in the BP parameter window on the patient
monitor display.
7. Disconnect the BP simulator cable from the BP connector of the
patient monitor.
8. Repeat steps for each enabled BP connector.
Use the following equipment for these tests:
GEMS-IT SpO
NOTE
The GEMS-IT SPO
GEMS-IT SPO
GEMS-IT SpO
1. Set the pulse oximetry (SpO
position.
2. Connect the Nellcor-style SpO
connector of the patient monitor and the SpO
Dash 3000/4000/5000
simulator (408610-001 or equivalent).
2
simulator is only compatible with the
2
parameter.
2
simulator cable (700232-004 or equivalent)
2
) simulator power switch to the off
2
simulator cable between the SpO
2
020A
2
simulator.
2
2000966-542D