Monitor defaults transfer
Storing monitor defaults
Copying stored monitor defaults
2000966-542D
Troubleshooting: Monitor defaults transfer
Below is a list of error messages that may display when performing a
patient monitor defaults transfer.
Message
ERROR - UNIT DEFAULTS HAVE NOT
BEEN SAVED
Message
ERROR COPYING UNIT DEFAULTS -
NETWORK ERROR
ERROR COPYING UNIT DEFAULTS -
UNIT TYPE MISMATCH
ERROR COPYING UNIT DEFAULTS -
SOFTWARE VERSION MISMATCH
ERROR COPYING UNIT DEFAULTS -
DATA CORRUPTION ERROR
ERROR COPYING UNIT DEFAULTS -
SOFTWARE FEATURE LEVEL
MISMATCH
ERROR COPYING UNIT DEFAULTS -
COUNTRY CODE MISMATCH
Dash 3000/4000/5000
Description
Patient monitor defaults were not saved on
the server patient monitor due to a memory
error in the patient monitor.
Description
Either an error has occurred while the
defaults were being transferred, or defaults
have not been saved on the server patient
monitor.
The server and client patient monitors are
set to different unit types (e.g., ADULT,
NEO, or OR)
The server and client patient monitors have
different software versions (e.g., V2C, V3A).
The patient monitor defaults transferred have
become corrupted, possibly due to a
memory error.
The server and client patient monitors are
set to different software levels (e.g., Basic,
Cardiac, Cardiopulmonary).
The server and client patient monitors are
set to different country codes (e.g.,
DEFAULT or FRANCE).