2000966-542D
Functional and electrical safety checks: Functional Checkout procedures
SpO
simulator configuration
2
1. Set the simulator as follows:
Set the MODE to NELLCOR.
Set the SpO
Set the PPR to 100 beats/min.
Turn the power ON.
Pulse oximetry test procedures
1. Verify that the following are displayed at the patient monitor:
A sinusoidal waveform with an SpO
An SpO
A PPR reading between 97 and 103 beats per minute (it might be
necessary to turn the SpO
2. Test the accuracy of these SPO
3. Test the accuracy of these PPR settings:
4. Return the simulator to these conditions:
Set the SpO
Set the PPR to 100 beats/min.
5. Press the NOISE TEST button on the simulator for 30 seconds.
6. Make sure the patient monitor still displays an SpO
97 and 100% (97 and 102%), or an interference detection message is
displayed.
7. Turn the simulator off.
8. Disconnect the simulator cable from the device under test.
Dash 3000/4000/5000
% to 99.
2
% reading between 97 – 100% (97 and 102%).
2
ON).
2
% settings.
2
Accuracy of SpO
Simulator setting
99% (Both types)
80.3% (84%)
49.7 (63%)
Accuracy of PPR settings
Simulator setting
70
100
160
% to 99.
2
label.
2
settings
2
Displayed SpO
value
2
97 – 100% (97 – 102%)
78 – 82% (81 – 87%)
48 – 52% (61 – 65%)
Displayed PPR value
68 – 72
97 – 103
156 – 164
value between
2
8-31