Recommended checkout

Recommended calibration, electrical safety tests, and checkout procedures
Replacement
Calibration
procedure
tests
Handle assembly
None
Display assembly
None
Display assembly
components
Display inverter
assembly
Speaker
None
Keypad assembly/
None
Trim Knob
7-52
Field replaceable units: Recommended checkout
After reassembling the patient monitor, ALWAYS complete the electrical
safety tests, checkout procedures, calibration tests, and regular
maintenance procedures identified and described in this section. Safety
and functional checkout tests should be documented in the checklist
provided in
"Checklist"
Checkout procedures
"Patient monitor power-up tests" on page
"Display test" on page
8-45.
"Wireless antenna signal strength test (option)"
on page
8-49. Use this procedure if the wireless
option is installed.
"Wireless LAN test (option)" on page
this procedure if the wireless option is installed.
"Patient monitor power-up tests" on page
"Display test" on page
8-45.
"Patient monitor power-up tests" on page
"Speaker test" on page
8-46.
"Wireless antenna signal strength test (option)"
on page
8-49. Use this procedure if the wireless
option is installed.
"Wireless LAN test (option)" on page
this procedure if the wireless option is installed.
"Patient monitor power-up tests" on page
"Display test" on page
8-45.
Dash 3000/4000/5000
on page D-1.
Electrical safety tests
8-21.
"Electrical safety tests" on page
8-52. Use
8-21.
8-21.
8-52. Use
8-21.
8-4.
2000966-542D