Field replaceable units: Remove or replace handle assembly
4. If your patient monitor has the wireless option, do the following to
replace the defective handle assembly:
CAUTION
Do not pull on the 802.11a/b/g antennas or use force
when connecting to the wireless card. Handle the
antennas and connectors carefully as they can easily be
damaged.
NOTE
Do not kink or bend the antenna cables tighter than a 5 mm
radius. Otherwise, the cable lining will delaminate and reduce
antenna performance.
a. Apply a piece of adhesive tape to temporarily hold the handle
onto the unit.
b. Remove the display assembly. Refer to
display assembly"
c.
To completely remove the handle from the main unit, do the
following:
Apply pressure against the antenna cables (to prevent
pulling on the cables) and carefully remove the antenna label
from the Dash monitor. Discard the used label.
Dash 3000/4000/5000
"Remove or replace
on page 7-18.