Bisx (Option) Current Leakage Tests; Bisx Patient (Source) Leakage Current Test - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

BISx (option) current leakage tests

BISx patient (source) leakage current test

8-16
Functional and electrical safety checks: BISx (option) current leakage tests
These procedures test the integrity of the BISx isolation only, not the
entire system.
This test checks leakage current from the patient cable connector of the
BISx to ground. Refer to the instructions contained with the safety
analyzer to perform each test.
NOTE
*The MD represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
Refer to the instructions contained with the safety analyzer to perform
this test.
1. Connect the Sensor Plus simulator to the BISx and connect the two
circular Sensor Plus simulator connections (test body) to the leakage
tester.
2. Configure the leakage tester as follows:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – CLOSED
Dash 3000/4000/5000
Sensor Plus simulator circular connections
502B
2000966-542D