7-38
Field replaceable units: Replace main unit parts
4. Remove the DAS assembly by first pulling the left side out 1/4-inch,
then sliding the whole assembly out of the patient monitor. If the
Wireless LAN option is installed, use a needle nose pliers to remove
the antennae by disconnecting both coax connectors from the
compact flash card.
5. Replace the DAS assembly.
6. Reassemble the unit reversing the order of removal steps.
7. If you installed the Generic Ohmeda DAS assembly, place the labels
as shown below.
NOTE
The BP 1 and BP 2 label looks similar to the example below.
Dash 3000/4000/5000
525A
BP 1 and BP 2
12SL (if software
option was enabled)
CO2, NBP, and ECG
861A
2000966-542D
862A