GE Dash 3000 Service Manual page 171

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

2000966-542D
Field replaceable units: Replace display flex assembly
6. Carefully bend the display flex as shown.
7. Pick up and support the display assembly between the main unit and
a thick book.
8. Use two screws to carefully secure the flex processor cable connector
to the main processor PCB.
9. Carefully align and insert the DAS cable pins into the DAS assembly
until the snap latch clicks into place.
10. Reassemble the patient monitor in reverse order of disassembly.
11. Complete the procedures in
Dash 3000/4000/5000
2 screws
"Recommended checkout"
874A
875A
876A
on page 7-52.