GE Dash 3000 Service Manual page 195

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

2000966-542D
Field replaceable units: Replace main unit parts
3.
Remove the assembly from the unit.
4. Align the tab on the power supply mounting bracket with the slot in
the rear housing and install the new assembly in the unit. Fasten the
assembly to the rear housing with the new screws.
NOTE
When securing the thumb screw on the writer flex, first finger-
tighten the thumb screw. Then, use a flat-head screw driver to
tighten the screw an additional 1/4 turn.
5. Reassemble the patient monitor in reverse order.
6. Complete the procedures in
Dash 3000/4000/5000
"Recommended checkout"
569A
slot
tab
576A
on page 7-52.
7-49