with display shield flex circuit
7. Secure the new keypad with the screws.
8. Make sure the washer tab fits in the
retaining slot of the Trim Knob control's
shaft and re-install the 11mm nut.
9. Secure the Trim Knob by reassembling
the display assembly in reverse order.
10. Complete the procedures in
"Recommended checkout"
52.

Replace display components without LCD

2000966-542D
Field replaceable units: Replace display assembly parts
Dash 4000 patient monitor
7. Reattach the keypad cable.
8. Secure the new keypad with the screws.
9. Make sure the washer tab fits in the
retaining slot of the Trim Knob control's
shaft and re-install the 11mm nut.
10. Secure the Trim Knob and reassemble
on page 7-
the display assembly in reverse order.
11. Complete the procedures in
"Recommended checkout"
52.
NOTE
These instructions are to replace the display components, not the
LCD. If the LCD fails, order the Display Assembly field replaceable
unit.
Dash 3000/4000/5000
Dash 5000 patient monitor
6. Secure the new keypad with the screws.
7. Secure the Trim Knob and reassemble
the display assembly in reverse order.
8. Complete the procedures in
"Recommended checkout"
52.
on page 7-
on page 7-
7-31