Pulse Oximetry Tests For Masimo Set Spo2 - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Pulse oximetry tests for Masimo SET SPO2

Equipment
Procedures
8-32
Functional and electrical safety checks: Functional Checkout procedures
Use the following equipment for these tests:
Masimo SpO
SPO2 Sensor Adapter Cable.
NOTE
The Masimo Tester is intended for use only with Masimo
instruments or instruments containing Masimo SET oximetry
with standard Masimo PC Series patient cable connectors. See
the Masimo "Directions for Use" that came with the Tester for
more information
The PPR and SPO2 values for the Masimo Tester must be within
these limits: Peripheral pulse rate: 61 bpm
value: 81%
1. Connect the Masimo-style SPO
connector of the patient monitor and the SPO
2. Turn on the patient monitor.
3. Verify the following are displayed at the patient monitor: (It might be
necessary to turn the SPO
A waveform with an SpO
An SPO
A PPR reading between 60 and 62 beats per minute.
4. Disconnect the simulator cable from the module.
Dash 3000/4000/5000
Test Kit (2021087-001). Includes Masimo Tester and
2

3%.
simulator cable between the SPO2
2
parameter on.)
2
label.
2
% reading between 78 - 84%.
2
1 bpm; and SPO2
simulator.
2
824C
2000966-542D

Hide quick links: