GE Dash 3000 Service Manual page 97

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Store monitor defaults for transfer
2000966-542D
Configuration: Advanced user procedures
To transfer monitor defaults, they must be set, then stored on the patient
monitor that is used as the server.
NOTE
To prevent the accidental erasure of the client patient monitor's
wireless LAN defaults, make sure that the server and client patient
monitors' wireless LAN defaults match before transferring patient
monitor defaults.
1. At the server patient monitor, set up the monitor defaults you want
to store. Refer to the following sections of the operator's manual for
reference:
Defaults Worksheet located in the preface of the manual. These
are blank and allow you to enter your own default settings.
Monitor Setup chapter.
Adult ICU, Neonatal ICU, Operating Mode Factory Default
Appendices.
2. When defaults are set, start from the Main Menu and select MORE
MENUS > MONITOR SETUP > SERVICE MODE.
3. Enter password using the Trim Knob control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407)
4. Select MONITOR SETTINGS.
5. Select STORE DEFAULTS FOR NETWORK TRANSFER.
6. Select YES to the confirmation popup menu.
Dash 3000/4000/5000
4-25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000Dash 5000001c 051d 003a

Table of Contents