GE Dash 3000 Service Manual page 228

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Patient simulator configuration
8-28
Functional and electrical safety checks: Functional Checkout procedures
Set up the patient simulator as follows:
Blood pressure (BP) polarity – POS, if required.
BP output – 0 mmHg.
BP transducer output – 5 µVolts.
1. Connect the BP simulator cable to enabled BP connectors you are
testing.
a. To connect to the BP1 or the BP2 connector, see the pictures
below.
Dash 3000/4000/5000
To Connect to the BP1 Connector
BP1 connector on monitor
blood pressure simulator
cable
patient simulator
863A
To Connect to the BP2 Connector
BP2 connector on monitor
blood pressure simulator
cable
patient simulator
864A
2000966-542D