Acceptance criteria NC
Acceptance criteria SFC – ground (earth), line or neutral open
2000966-542D
Functional and electrical safety checks: Electrical safety tests
13. Read and record the current leakage indicated on the tester.
14. Set the power switch of the device under test to OFF.
If measured reading is greater than the appropriate specification below,
the device under test fails. Contact GE Technical Support.
100 microamperes (0.1 volts on the tester), and the device under
test is powered from 100-240 V/50-60 Hz
(USA only) 300 µA, and the device under test is powered from
100-120 V/50-60 Hz
(USA only) 300 µA, and the device under test is powered from a
center-tapped 200-240 V/50-60 Hz, single phase circuit (UL Split
Phase Exemption)
500 µA, and the device under test is powered from a non-center-
tapped, 200-240 V/50-60 Hz, single-phase circuit
NOTE
If the reading is greater than the specification below, and the
device under test is powered from 100-240 V/50-60 Hz, the device
under test fails. Contact GE Technical Support.
Dash 3000/4000/5000
8-11