Severity
Date and time
Error number
2000966-542D
Troubleshooting: Review errors
70F
Date was changed from this patient monitor. This helps determine
how the system-wide date was altered.
710
Incompatible software was detected. If the main processor software
finds that the software operating on the communication software
incompatible, it turns off the communication (network) controller and
enters this data into the error log. When the patient monitor won't
"talk" to the network, looking for this entry in the error log is one part of
the troubleshooting process.
NOTE
The patient monitor may be referred to as a display or scope in the
error code descriptions.
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Continue—the event/error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log entries
will have a severity of Continue.
Fatal—the event/error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a WARM
START.
Forced Restart—the system was restarted by a known condition
(internet address change, video test, etc).
The date and time the event or problem occurred.
A sequential number that is used to identify each event/problem.
Dash 3000/4000/5000
Definition of error codes
6-15