Replace CPU/battery housing assembly

7-44
Field replaceable units: Replace main unit parts
1. For Dash 3000 patient monitors, remove the handle assembly. Refer
to page 7-10. For Dash 4000 or 5000 patient monitors, go to the next
step.
2. Remove the display assembly from the main unit. Refer to page 7-18.
3. If present, remove the wireless card.
CAUTION
Do not pull on the 802.11a/b/g antennas or use them to
remove the wireless card. They can easily be damaged.
a. Release the antenna cables from underneath the left side of the
antenna label to allow enough slack in the cable to allow the
wireless card to be removed from the PC card bracket.
Dash 3000/4000/5000
961B
2000966-542D