Boot Code Settings; Set Defib Sync Voltage And Pulse Width; Set Line Frequency - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Boot Code settings

Set Defib Sync Voltage and pulse width

Set Line Frequency

4-18
Configuration: Boot Code settings
To activate the Boot Code,
1. With power on, hold down NBP Go/Stop (or NBP Auto on Dash
5000) and ZERO ALL.
2. Press and release the Trim Knob control.
3. Keep holding NBP Go/Stop (or NBP Auto) and ZERO ALL until
the Boot Code information appears on the display.
The patient monitor controls the analog out signal used to trigger a
defibrillator. Refer to the defibrillator manufacturer's manual for the
required pulse amplitude and duration.
Configure or change the Marker Out signal of the DEFIB SYNC
connector as follows.
1. Activate Boot Code
SET CONFIGURATION menu option.
2. In the CONFIGURATION MENU, select:
1 Defib Sync Voltage: and choose 5V or 12V amplitude.
2 Defib Sync Pulse Width: and choose 10 ms or 100 ms for
pulse duration.
3. Select Exit and wait for the patient monitor to display the message,
"Writing settings to Flash config mem..."
4. Manually reboot the patient monitor to implement this change.
Configure or change the patient monitor line frequency to 50 or 60 Hz as
follows. The default is 60 Hz.
1. Activate Boot Code
SET CONFIGURATION menu option.
2. Select Line Frequency then choose 50 Hz or 60 Hz line frequency.
3. Select Exit and wait for the patient monitor to display the message,
"Writing settings to Flash config mem..."
4. Manually reboot the patient monitor to implement this change.
Dash 3000/4000/5000
("Boot Code settings" on page
("Boot Code settings" on page
4-18) and select
4-18) and select
2000966-542D