BP connector test procedures
2000966-542D
Functional and electrical safety checks: Functional Checkout procedures
b. To connect to the BP3 or BP4 connectors, see the pictures below.
BP1
BP3
BP2
BP4
1. Verify the BP parameter window, waveform label, corresponding
graticules, and waveform appear on the patient monitor display,
along with a BP waveform requiring zero reference.
2. Press the ZERO ALL (FUNCTION) key on the front panel of the
patient monitor to zero-reference the BP waveform.
Dash 3000/4000/5000
To connect to the BP3 connector
BP1/BP3 connector on monitor
dual BP cable,
PN 2005772-001
866A
blood pressure simulator
cable
To connect to the BP4 connector
BP2/BP4 connector on monitor
dual BP cable,
PN 2005772-001
blood pressure simulator
867A
cable
patient simulator
patient simulator
8-29