5 Operation
A-START/B-START
(starting point of the gate)
You can set the starting point of the gates A or B within
an adjustment range of 0 ... 27940 mm.
You can set the value by coarse adjustment steps or by
fine adjustment (see Section Coarse adjustment and
fine adjustment, page 4-11).
– Select the function A-START or B-START.
– Set the required value.
5-22
Setting the gates (function groups GATE A and GATE B)

A-WIDTH/B-WIDTH

(width of the gates)

You can set the width of the gates A or B within an ad-
justment range of 1.00 ... 27940.00 mm.
You can set the value by coarse adjustment steps or by
fine adjustment (see Section Coarse adjustment and
fine adjustment, page 4-11).
– Select the function A-WIDTH or B-WIDTH.
– Set the required value.
Issue 2 (12/2013)
USM 36