Storing The A-Scan And Parameters In The Test Report - GE USM 36 Series Technical Reference And Operating Manual

Ultrasonic flaw detector
Table of Contents

Advertisement

Test reports
Storing the A-scan and parameters in the
test report
When storing test reports, you can decide whether to
store the A-scan and the list of parameters as well. To
do this, you have to turn on the corresponding function.
USM 36
– Switch to the second operating level.
– In the function group FILES, select the function
IMAGE IN REPOR and choose YES.
– Select the function PARAM IN REPOR and choose
YES.
Next time a test report is stored, the selected informa-
tion is inserted.
Issue 2 (12/2013)