5 Operation

Sensitivity correction

You can use the function TRANSFER CORR. to com-
pensate for transfer losses in the material to be tested.
A correction is necessary if the test object and the refer-
ence standard have different surface qualities.
You have to determine the adjustment value for the
compensation of transfer losses by way of experiment.
In this connection, the gain is varied accordingly but the
curve shape remains the same.
– Select the function group MAT ATTN.
– Select the function TRANSFER CORR. and set the
required value.
5-108
Distance-amplitude curve according to JIS Z3060-2002 (JISDAC)

Turning the JISDAC evaluation off

You can turn the JISDAC evaluation off at any time.
– Switch to the first operating level.
– Select the function group JISDAC.
– Press one of the selector keys to display the specific
function groups and functions for the JISDAC.
– Select the function group SETUP.
– Select the function JISDAC and press one of the se-
lector keys to turn the JISDAC evaluation off (OFF).
The DAC curve and the evaluation lines are no longer
displayed in the A-scan.
N o t e
The DAC curve is not lost by turning the func-
tion off. By turning it on again, you can use
the function JISDAC to go back to the JIS-
DAC evaluation again without losing any set-
tings.