BLOCK

You can read the preselected test ranges of various cal-
ibration standards in the function group BLOCK.
Example: K2 65-75
The side-drilled hole is located at a depth of 7.68 mm
and has a diameter of 5.00 mm.
5-50
The option CUSTOM in the function BLOCK (function
group AUTOANG) enables you to design a calibration
standard of your own and to enter its values in the func-
tions DIAMETER and DEPTH (function group BLOCK).
In this regard, it is important that DEPTH always refers
to the center of the side-drilled hole and not to the real
reflecting surface.
Issue 2 (12/2013)
Defining the probe angle