General setup
TOF in LAYER
As an alternative to the display of an exact value (time-
of-flight or distance), the USM 36 can only display the
layer in which a reflector is located.
You can specify layers of varying depth (but with identi-
cal material velocity) for this purpose. They are auto-
matically numbered consecutively. With the function
TOF in LAYER turned on, only the number of the layer
is displayed in which the flaw (e.g. inclusion or separa-
tion) is located.
USM 36
Defining layers
You can define either the thickness of a single layer or
the thicknesses of 10 layers within a range of
1.00 ... 10.0 mm each.
– Switch to the second operating level.
– Select the function group CONFIG4.
– Select the function TOF in LAYER and press one of
the selector keys to turn on the function.
– Switch to the function LAYER TYPE.
– Select the mode for a single layer (STANDARD) or
for 10 layers (CUSTOM).
– Switch to the function LAYER EDIT and press one of
the selector keys. The list of layers with the corre-
sponding thicknesses in each case is displayed.
– Set the required values.
– Press one of the selector keys to complete the set-
ting.
This completes the definition of the layer thicknesses.
Issue 2 (12/2013)
5 Operation
5-81