Saving the settings

Recalling settings

You can recall and use instrument settings saved to the
SD memory card.
N o t e
You can only recall datasets having the file-
name extension UGO. Other files on the SD
memory card are not displayed as options.
– Switch to the second operating level.
– In the function group FILES, select the function
DIRECTORY and press one of the two selector keys.
The directory of the SD memory card appears.
– Press the arrow keys to select a directory on the SD
card.
– Press one of the two selector keys to choose the se-
lected directory.
USM 36
– Switch to the function ACTION.
– Select the function RECALL DATASET.
– Switch to the function FILENAME.
– Select the name of the required file. The selection
switches automatically to the function ENTER.
– Press one of the two selector keys.
The instrument settings of the selected dataset are re-
called and they are immediately active after closing the
recalling process.
Issue 2 (12/2013)
4 Principles of operation
4-21