5 Operation
A T T E N T I O N
Use the data sheet for your probe to check
which maximum voltage is allowed to be ap-
plied.
N o t e
The pulser voltage and the pulse width can
be automatically limited, depending on the
pulse repetition frequency (see Section PRF
MODE (pulse repetition frequency),
page 5-16) setting. This function helps to
avoid heat accumulations in the pulser elec-
tronics.
– Select the function VOLTAGE.
– Select the required pulser voltage.
5-14

ENERGY

You can use the function ENERGY to set the penetra-
tion or sound energy.
The following settings are possible:
● HIGH – high energy
● LOW – low energy
The setting HIGH is recommended for all tests in which
maximum sensitivity is important, e.g. for the detection
of small flaws. Choose the setting LOW for broadband
probes or if narrow echoes are required (better lateral
resolution).
– Select the function ENERGY.
– Select the required setting.
Issue 2 (12/2013)
Setting the pulser (function group PULSER)
USM 36