Calibration Reminder - GE USM 36 Series Technical Reference And Operating Manual

Ultrasonic flaw detector
Table of Contents

Advertisement

General setup

Calibration reminder

Certain test specifications require a recalibration of the
entire system (instrument, cable, probe, material) at
regular time intervals.
If you turn the reminder function of the USM 36 on, a re-
minder icon is displayed next to the A-scan as soon as
the scheduled point of time is reached (see Section Sta-
tus display icons, page 0-7, at the beginning of this op-
erating manual).
USM 36
You can turn on a short-term calibration reminder for a
time period within the range of 0.5 ... 4 hours and a re-
minder for the yearly calibration.
In the case of the yearly reminder, define the date at
which the calibration should be carried out. You can de-
cide how long before this date the reminder icon should
be displayed.
Short-term calibration reminder
– Switch to the second operating level.
– Select the function group CONFIG3.
– Select the function CAL REMINDER and select the
required period of time.
– Switch to the function CAL RESET and press one of
the selector keys to acknowledge the reminder and to
reset the starting point for the defined period of time
to zero.
Issue 2 (12/2013)
5 Operation
5-87