Chapter 12. CALIBRATION
PV zero, span
Writing into EEPROM
12-12
(1) PV zero adjustment
(a) Press SETUP key until the PROG/SEG display shows (
(b) Adjust your calibration device to an output signal equal to the 0% range value
(See Table 12-9), the signal need to be on the input for 10 to 15 seconds.
(c) Press ENTER key after display stabilizes.
(2) PV span adjustment
(a) Press SETUP key until the PROG/SEG display shows (
(b) Adjust your calibration device to an output signal equal to the 100% range value
(See Table 12-9).
(c) Press ENTER key after display stabilizes.
Gain No.
PV Input 0% PV Input 100%
0
Ð5.000mV
1
Ð10.000mV
2
0.000mV
3
0.000mV
4
0.000mV
5
Ð1.000V
6
0.000V
7
0.000V
8
4.000mA
9
9
20
9
10
20
9
11
1
12
0.000V
*
Note:
* ; If model isnÕt Carbon Potential, it isnÕt necessary to calibrate this item.
Press SETUP key until the PROG/SEG display shows (
Scroll ↑, ↓, ←, or → keys to show (
Press ENTER key.
Table 12-9.
Connecting Position
10.000mV
Between A(+) and B(Ð)
18.000mV
Between A(+) and B(Ð)
36.000mV
Between A(+) and B(Ð)
52.000mV
Between A(+) and B(Ð)
100.000mV
Between A(+) and B(Ð)
1.000V
Between A(+) and B(Ð)
5.000V
Between A(+) and B(Ð)
10.000V
Between A(+) and B(Ð)
20.000mA
Between C(+) and B(Ð)
9
140
Between A(+) and B(Ð)
9
480
Between A(+) and B(Ð)
9
100
Between C(+) and B(Ð)
1.250V
Between C(+) and B(Ð)
1.1.1.1.
) on the upper display.
01-03
).
01-04
).
Figure
Fig.12-5
Fig.12-8
Fig.12-9
Fig.12-6
Fig.12-7
Fig.12-10