Changing the recorder and printer settings

Recorder settings

1.
Press
2.
Select Record Waveforms.
3.
Select Record Trends.

Printer settings

1.
Press
2.
Select Print Graphical.
.
Print/Record
Select Waveform 1 and select a parameter or select OFF. Then select Waveform 2
and 3 and their parameters. You can record up to three waveforms simultaneously.
If you wish to record only one waveform, select the other waveform fields OFF.
Select Start on Alarms - Yes to start automatic strip chart recording when the
following alarms reach the red alarm level (NO is the default):
with L-DICU08 software license: bradycardia, tachycardia, asystole, Art high/
low, V Fib, V Tachy
Select Delay and OFF or 12 s. If the recording delay is OFF, the recording starts
when an event occurs and continues for 30 seconds or until manually stopped, or
until the recorder runs out of paper. If the delay is 12 seconds (default), the recording
starts when an event occurs and the 12 seconds prior to the event are recorded
from the recorder memory. The recording continues for 18 seconds if the length has
been set to 30 seconds or until the recorder runs out of paper.
Select Paper Speed and 1, 6.25, 12.5 (default), or 25 mm/s.
Select Length and choose 30 s (default) or Cont.
Select Trend Resolution and every 1 min, 5 min, 10 min (default) or 30 min. This
setting is for numerical trends.
Select Num Trend Type and Num. (default) or Tab. as the format of the numerical
trend recorded.
Select Graphic. Trend 1 and select the parameter and then do the same for Graphic
Trend 2. These settings define the graphical trends recorded in upper field and lower
field. You can record graphical trends of two parameters.
.
Print/Record
Hours/Page and select to print 1, 2 (default), 4, 6, 8, 10, 12, 24, 36, 48 or 72 hours on
one page.
Select Trend Length and select 1, 2, 4, 6, 8 (default), 10, 12, 24, 36, 48 or 72 hours.
Monitor setup