Regular Checks - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor

Regular checks

When you start monitoring, check that:
The module is firmly in place.
Accessories are intact and properly connected.
You have selected desired parameters to be displayed in the digit and waveform fields.
ECG and impedance respiration
Check that the message 'Leads off' disappears and the waveforms are displayed when
the cable is connected to the patient.
Pulse oximetry
Check that the red light is lit in the sensor.
Check that the SpO
when the sensor is connected to the patient.
Temperature
Check that the temperature value is displayed when the probe is connected to the
patient.
InvBP
Check that the monitor recognizes cable connections (activates the display) for all the
pressure channels used and the pressure values are shown.
Make sure all transducers are zeroed correctly.
NIBP
Ensure that you are using correct cuff size and have selected correct inflation limits. For
children and when using hoses without identification, the inflation limit must be set
manually.
Check that the cuff hose detection works properly.
Check that the pressure values are displayed.
Start the Venous Stasis mode and check that the pump is not restarting during the
measurement. If it does, the cuff may be leaking.
Airway gas (CO
Check that the water trap is empty.
Occlude the sampling line and check that the message 'Sample line blocked' appears
within 30 seconds and gas waveforms are showing zero at the same time.
Functioning of the alarms
Set a parameter value outside the alarm limits. For example, connect the SpO
adjust the SpO
priority (Yellow) to high priority (red) according to sequence given in the alarm categories
table, see section
Check that the yellow and red LEDs function as indicated in the table. To check audible
alarms and alarm light, see section
example because it is often off by default, it is easy to use for alarm checking.
If the monitor does not work as described, see section
9-2
value is displayed and the message 'SpO
2
)
2
High limit under the measured SpO
2
"Alarms."
"Alarms."
probe off' disappears
2
values. The alarms go from medium
2
NOTE: Although SpO
may not be the best
2
"Troubleshooting."
sensor and
2