B30 Patient Monitor
Message
Lead changed
Leads off
Network down: xxx
NIBP manual
No battery backup
Noise
No P1 transducer
No SpO2 probe
No SpO2 pulse
Poor signal
Printer error
Printing...
Printing ready
PVC rate limit changed
10-4
Explanation
The monitor automatically switches
the ECG1 waveform selection to a
measurable ECG Lead (I, II, III, aVR, aVL,
aVF or V) if the current ECG1 is not
measurable. Note that the ECG
waveform changes according to the
lead it is measured from.
ECG trunk cable or any of the leads is
disconnected. Offset voltage between
two electrodes is too high.
Network cable is not connected.
The Central is shut down.
Autocycling mode is interrupted
because of an air leak or loose cuff.
Unreliable HR calculation or distorted
waveform, possibly during
defibrillation or because of motion
artifacts.
Transducer or channel x cable
disconnected.
There is no SpO
probe.
2
No SpO
pulse can be detected.
2
Signal is not sufficient.
Printer is not working properly.
Printing network printer has started.
Monitor has completed sending
printing data to the printer.
Limit has changed on the Arrhythmia
Workstation
What to do
Check the lead.
ECG: Reconnect the disconnected
trunk cable, electrode or leadwire.
ECG: Change the trunk cable,
leadset and module.
Check the network cable.
Check the Central.
Check the NIBP setup and restart
autocycling.
Replace the batteries.
ECG: Check the patient status.
ECG: Check the electrodes.
Ensure that the patient is not
shivering.
Poor electrode quality or wrong
positioning.
Change the lead.
Remove the ECG cable from the
connector and reinsert it.
Connect the transducer or the
cable.
Check connection between the
SpO
sensor and module.
2
Try other measuring sites.
SpO
: Change the measuring site.
2
Check that the network printer is
operational.
Please wait until the printing is
finished.
Please wait until the printing is
finished.
Check the limit.