5 Monitor setup

Overview

The B30 monitor has numerous setup options for screen, parameters, alarms, etc. There are
two types of settings:
User mode settings: Some of the monitor setup options are preconfigured to be effective.
The preconfigured default settings, so called factory settings, form six sets of user modes.
The monitor starts in the startup mode, which is one of the user modes. You can change
to another user mode, see
The changes you make in the user mode settings are valid only temporarily until you
discharge the patient or change a mode, or until more than 15 minutes has elapsed from
the turn-off of the monitor. The changes need to be saved in the mode to become
permanent.
Monitor installation settings: the monitor installation settings are the same in all user
modes. The changes are permanent and preserved until changed again.
Before starting to use the monitor, check the monitor installation settings and what is
configured in the different user modes, and make necessary changes. You need passwords for
making the changes, see
This User's Reference Manual describes most of the configurable features. More detailed lists of
the options, and also of the factory settings, can be found in the "Default Configuration
Worksheet" delivered with each monitor.
NOTE: If you wish to make permanent changes, we recommend that you contact the person
responsible for the configuration, who is familiar with the configuration architecture. When
new settings are saved, they should be marked in the "Default Configuration Worksheet."

Configuration and passwords

NOTE: If you want to make changes that require a password, we recommend you contact the
system administrator.
The default password for entering the Install/Service menu is 16, 4, 34.
Monitor Setup
Press
adjustment window until you hit the desired number, then push the ComWheel to accept and
select the number. Continue until all three numbers are selected. After entering the third
number the Install/Service menu is displayed.
Most of the modifications are valid only temporarily unless you accept them in the Save Modes
menu, which is a submenu of the Install/Service menu. A password is also required for
entering the Save Modes menu.
The default password for entering the Save Modes menu is 13, 20, 31.
"Changing the user mode" page 5-6.
"Configuration and passwords"
and select the Install/Service menu, turn the ComWheel in the opened