Using Venous Stasis - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor

Using venous stasis

Venous Stasis initiates and holds the pressure in the cuff to help venous cannulation. A
constant pressure is maintained in the cuff. The value is selected automatically after the
inflation limits (infant, child or adult) are set.
NIBP
To release the pressure before two minutes have expired:
Press the
press the
The inflation limits, venous stasis pressure and venous stasis times are listed in the table below.
Inflation limit
Infant
150+5 mmHg (19.3 ± 0.7 kPa)
Child
200+10mmHg (26.7 ± 1.3 kPa)
Adult
280+10 mmHg (37.3 ± 1.3 kPa)
Venous stasis pressure may be lower than the values above if the patient has low blood
pressure. The venous stasis pressure adapts to the measured mean pressure being
approximately the same as the mean pressure but always at least the following:
Infant 20 ± 5 mmHg (2.7 ±0.7 kPa)
Child 30 ± 5 mmHg (4.0 ±0.7 kPa)
Adult 40 ± 5 mmHg (5.3 ±0.7 kPa)
16-8
Start/Cancel
key on the module, or
NIBP
key on the Command Board and select Stop Ven. Stasis.
Maximum venous stasis
inflation limit
40+5mmHg (5.3 ± 0.7 kPa)
60+5mmHg (8.0 ± 0.7 kPa)
80+5 mmHg (10.7 ± 0.7 kPa)
Venous stasis time
1 min
2 min
2 min

Advertisement

Table of Contents
loading

Table of Contents