B30 Patient Monitor
Recording time is 30 seconds consisting of 12 seconds recording from the recorder memory
and 18 seconds real-time recording. The alarm source is always marked to the alarm
recordings. The following alarms start the recording:
Alarm
Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy
V Run >3
Alarming level
Severe
Showing alarm history
•
Press the
last 20 alarms that have reached the yellow or the red alarm level. The time and type of
occurrence are displayed next to the alarm list items.
4-12
Recorded parameters
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
Alarms which start the recording
Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy
Pt. Data & Trends
key and select
Alarm History
. This displays a list of the