Recording on alarms

An automatic strip chart recording is activated when the following alarms reach the red level:
Asystole, Tachy, Brady, Art High, Art Low, V Fib, and V Tachy. When recording is activated by
alarms, the recording time is always 30 seconds and the delay always 12 seconds.
1.
Press the
2.
Select Record Waveforms.
3.
Select Start on Alarms and Yes.
Print/
Record
The following alarms start recording:
Alarm
Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy
Arrhythmia analysis
Severe
All alarm recordings are marked with the alarm source.
Print/Record
key.
Alarms which start the recording (if the alarm
is set on/level red)
Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy
Recorded parameters
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
Printing and recording
8-5