GE Healthcare B30 User's Reference Manual page 153

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

Table 10-4
Other problems related to batteries
Symptom
Possible cause and solution
Battery operation
time is markedly
shortened
Table 10-5
Other problems related to ECG measurement
Symptom
Possible cause and solution
ECG signal is
noisy or no QRS is
Incorrect ECG filter.
detected
Poor electrode quality or wrong positioning.
Pacer markers are
not visible
-
-
-
Thick ECG
ECG cable is looped.
baseline
Other electrical power cables are near the ECG leadwires.
Incorrect ECG filter.
Incorrect power frequency of the monitor.
Table 10-6
Other problems related to impedance respiration measurement
Symptom
Possible cause and solution
Respiration
measurement
fails
Table 10-7
Other problems related to InvBP measurement
Symptom
Possible cause and solution
InvBP readings
seem unstable
Condition the batteries according to the instructions in this
manual.
Ensure that the patient is not shivering.
Check the filter through
Check the electrodes and cables and their placement. See
section"ECG"
for details.
Change the lead.
Remove the ECG cable from the connector and reinsert it.
Check that:
The pacer markers have been selected ON.
The pacemaker has been adjusted correctly and not above R.
The pacemaker functions correctly: ECG cables, electrodes and
setups are correct.
Check the electrode quality and positioning.
Adjust the detection limits. During ventilator-supported breathing, the
respiration calculation may count only ventilator-produced
inspirations and expirations.
Other electrical devices may interfere with the measurement.
Make sure there are no air bubbles in the transducer system. Flush
and zero.
Place the transducer on the patient's mid-heart level and zero.
ECG
- ECG Setup - Filter.
Troubleshooting
10-7