GE Healthcare B30 User's Reference Manual page 58

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

B30 Patient Monitor
Direct current for leads-off detection
through a reference electrode:
The normalized respiration sensing current between RA (R) and LL (F)
or RA (R) and LA (L) or LA (L) and LL (L):
Frequency of respiration sensing current:
Minimizing the effects of the line isolation monitor transients:
Crystal controlled oscillator used as the operating frequency source of the patient isolation power
supply.
The average time and time range ( ) to alarm (VFib or VTachy) for tachycardia waveform
are as follows (IEC60601-2-27 6.8.2.bb.6):
Figure 101 B1 halved amplitude:
Figure 101 B1 normal amplitude:
Figure 101 B1 doubled amplitude:
Figure 101 B2 halved amplitude:
Figure 101 B2 normal amplitude:
Figure 101 B2 doubled amplitude:
A clinician should always confirm the rhythm from the ECG waveform.
Direct cardiac application
The display area reserved for the ECG measurement in the screen may not be adequate for displaying
the complete ECG amplitude when measuring ECG direct from the surface of the heart. Clipping of the
signal can be reduced by adjusting the size of the signal on the screen (for example, from the default 1.0
to 0.2) in the ECG menu.
Impedance respiration
Respiration range:
Accuracy:
The conducted RF immunity of the respiration measurement has been tested with 1 Vrms. The radiated
RF immunity of the respiration measurement has been tested with 1 V/m. The impedance respiration
measurement technology has been optimized so that the measurement is not sensitive to electrosurgery
equipment which is commonly used in the intended environment for the impedance respiration
measurement.
NOTE: Impedance respiration measurement is intended for patients ages three years
and up.
Invasive blood pressure
Measurement range:
Measurement accuracy:
Pulse rate:
Measurement range:
Accuracy:
Transducer sensitivity:
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a
defect in the ESU return electrode connection.
2-32
9.9 s (8.4 to 11.5 s)
7.1 s (5.8 to 8.2 s)
4.4 s (4.2 to 4.6 s)
7.0 s (6.1 to 7.5 s)
5.8 s (4.5 to 7.4 s)
6.1 s (5.1 to 7.0 s)
4 to 120 resp/min
± 5% or ± 5 resp/min, whichever is greater
1
(E-PSMP module only)
-40 to 320 mmHg (-5.3 to 42.6 kPa)
±5 % or ±2 mmHg
30 to 250 bpm
±5 % or ±5 bpm
5 µV/V/mmHg
<120 nA
<5.0 µA
31.25 kHz

Advertisement

Table of Contents
loading

Table of Contents