B30 Patient Monitor
Module description
E-PSMW
Figure 15-1 Modules E-PSMW and E-PSMPW for impedance respiration
Respiration detection
The respiration rate is the sum of the respirations that have exceeded the detection limit.
The dotted lines present the zero line and the detection limit. The signal strength produced by a
respiration should thus exceed this minimum limit to be included in the respiration rate
calculation. Peaks within the grids are not calculated.
%
If the detection mode is AUTO, the grid lines present the minimum limits. The limits in use may
be larger. The RR value could include fewer respirations than indicated by the gridline.
Respiration rate calculation
Respiration rate is calculated automatically when ECG or CO
respiration measurement is turned off. When CO
automatically calculated from CO
Respiration rate calculation switches back to impedance respiration if you press the
Alarms
The respiration rate source is displayed above the respiration numeric value.
Figure 15-2 Respiration rate source indication for impedance and CO
The impedance respiration waveform is displayed next to the RR value also when the RR value
is calculated from CO
15-2
ECG and
impedance
respiration
connector
measurement
key during an Apnea alarm.
.
2
E-PSMPW
is measured, the respiration rate is
2
.
2
is measured unless the
2
2
Silence